P70 piezo nanopositioning scanner has two versions, XY-motion and Z-motion version. Both versions have four through holes at the four corners, and the hole center distance is 30mm, which is very suitable for the installation of cage structures.
|
Moving direction
|
Applied to Cage Structures
|
Applied to Cage Structures
|
| Technical data | |||
|
Models |
P70.Z8S |
Units |
|
|
Active axes |
Z |
|
|
|
Drive control |
1 driving channel |
|
|
|
Travel |
6.4(@120V) |
μm±10% |
|
|
Max.travel |
8(@150V) |
μm±10% |
|
|
Sensor type |
SGS |
|
|
|
Resolution |
0.3/0.1 |
nm |
|
|
Linearity |
0.25/- |
%F.S. |
|
|
Repeatability |
0.16/- |
%F.S. |
|
|
Pitch/yaw/roll |
<10 |
μrad |
|
|
Load capacity |
Horizontal |
0.2 |
kg |
|
Vertical |
0.03 | ||
|
Inverted |
0.05 | ||
|
Response time |
20 |
ms±20% |
|
|
Unloaded resonance |
3.6 |
kHz±20% |
|
|
Load resonance |
1.5(200g) |
kHz±20% |
|
|
El.capacitance |
1.6 |
μF±20% |
|
|
Mass |
128.5 |
g±5% |
|
Note: Max driving voltage could be -20V~150V, Recommended voltage 0~120V for long-term operation to extend lifetime.
Technical data are measured by CoreMorrow E00/E01 series piezo controller.
| Mounting Method |
Horizontal Mounting |
Vertical Mounting |
Inverted Mounting |
| Definition |
Place the PZT platform horizontally for installation, with the moving surface located above the PZT platform and parallel to the horizontal plane.
PS: Default direction for carrying capacity in the parameter table
|
Place and install the PZT platform vertically, with the moving surface parallel to the direction of gravity
|
Place the PZT platform horizontally, but the moving surface is located below the PZT platform |
| Diagram |
![]() |
![]() |
![]() |
| Note |
When the mounting method is selected as vertical mounting, pay attention to the outlet port |
||
|
|
||||||||||||