Technical Data
|
|
|
|
Technical data | ||||
Type |
P77A.100S75 |
P77A.100K75 |
Units |
|
Active axes |
Z |
Z |
|
|
Sensor type |
SGS | - |
|
|
Travel range(0~120V) |
80 |
80 |
μm±10% |
|
Max.travel range(0~150V) |
100 |
100 |
μm±10% |
|
Aperture |
Φ75 |
Φ75 |
mm |
|
Length/OD |
Φ145 |
Φ145 |
mm |
|
Height |
33 |
33 |
mm |
|
Resolution |
3 |
1 |
nm |
|
Load capacity |
Horizontal |
1 |
1 |
kg |
Vertical |
1 | 1 | ||
Inverted |
3 | 3 | ||
Electrical capacitance |
19.2 |
19.2 |
μF±20% |
|
Mass |
1260 |
1260 |
g±5% |
|
Operating temperature |
-20~80 |
-20~80 |
℃ |
|
Control |
2channel |
2channel |
|
|
Connector |
LEMO |
LEMO |
|
|
Cable length |
1.5 |
1.5 |
m±10mm |
|
Material |
Al, Steel |
Al, Steel |
|
Note: Max driving voltage could be -20V~150V, Recommended voltage 0~120V for long-term operation to extend lifetime.
Technical data are measured by CoreMorrow E00/E01 series piezo controller.
Mounting Method |
Horizontal Mounting |
Vertical Mounting |
Inverted Mounting |
Definition |
Place the PZT platform horizontally for installation, with the moving surface located above the PZT platform and parallel to the horizontal plane.
PS: Default direction for carrying capacity in the parameter table
|
Place and install the PZT platform vertically, with the moving surface parallel to the direction of gravity
|
Place the PZT platform horizontally, but the moving surface is located below the PZT platform |
Diagram |
![]() |
![]() |
![]() |
Note |
When the mounting method is selected as vertical mounting, pay attention to the outlet port |
|
|